Functional Tests Installation Guide
2 - 30 046-001141-00 A5/A3â„¢ Service Manual
FIGURE 2-42 O2 Sensor Calibration Canceled
4. The screen shown below is displayed if the calibration has failed. A Fail code is displayed in red.
Select Try Again to repeat the calibration. Select Done to exit the calibration screen.
FIGURE 2-43 O2 Sensor Calibration Failed
5. The screen shown below is displayed after a successful calibration. Select Done to exit the
calibration screen.