A5/A3â„¢ Service Manual 046-001141-00 3 - 21
Periodic Maintenance Functional Tests
FIGURE 3-20 Calibration Front Control Panel - Flow Setting
c. Gas Settings: Press the Setup button, select Setting->ENTER->Gas Settings->MODIFY->Gas
Type->O2.
d. Correction mode: Press the Setup button, select Setting->ENTER->Correction Mode-
>MODIFY->Correction Mode->ATP.
FIGURE 3-21 Calibration Front Control Panel - Gas Settings
e. Select BACK->BACK-> BACK.