EasyManuals Logo

Mindray A3 Service Manual

Mindray A3
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
Functional Tests Periodic Maintenance
3 - 22 046-001141-00 A5/A3â„¢ Service Manual
FIGURE 3-22 Calibration Front Control Panel - High Flow
Calibration Device (TSI Certifier 4070)
a. Setup the calibration device as described below.
b. Flow Setting: Press the to line select key until the top display reads LPM or SLPM. If the
display reads SLPM, press the DISPLAY UNITS key until the top display reads LPM.
FIGURE 3-23 TSI Certifier 4070 Menu
c. Gas Settings: Press the GAS SELECT key until the bottom left corner reads O2.
6. When the system is in Standby, select the Setup-> Service-> Diagnostic Tests->Valves to access
the following menu.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A3 and is the answer not in the manual?

Mindray A3 Specifications

General IconGeneral
BrandMindray
ModelA3
CategoryMedical Equipment
LanguageEnglish

Related product manuals