A5/A3â„¢ Service Manual 046-001141-00 3 - 25
Periodic Maintenance Functional Tests
FIGURE 3-27 Calibration Device (VT Plus)
FIGURE 3-28 TSI Certifier 4070
2. Set up the calibration device as described below.
To set the Fluke VT-Plus:
a. Pressure Setting: Press the PRESSURE button on the front control panel of the calibration,
select Range and then set it to Low Press.