EasyManua.ls Logo

Mindray A3 - Page 193

Mindray A3
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A5/A3™ Service Manual 046-001141-00 4 - 7
Calibration System Calibration
4.3.1.1 21% O2 Calibration
Follow these steps to calibrate O2 sensor at 21% O2.
1. Select Setup > General > Calibrate O2 Sensor or Setup > System > Calibration> O2 Sensor or
Setup > Service > Calibration > O2 Sensor to access the screen as shown below. The General tab
shows only 21% O2 Sensor calibration; the System and Service tabs require passwords and show
both 21% and 100% O2 Sensor calibration. Set up the machine as per the instructions on the
screen. Select Begin to start calibration.
FIGURE 4-1
2. The calibration screen shown below is displayed when Begin is selected. During the calibration,
you can select Cancel to cancel the calibration.

Table of Contents

Related product manuals