A5/A3â„¢ Service Manual 046-001141-00 4 - 57
Calibration System Calibration
4.3.5.1 Zeroing Procedures
Follow these steps to zero pressure and flow sensors.
1. Select Setup-> Service-> Calibration-> Zero Sensors to access the screen shown below.
FIGURE 4-69
2. Select Begin to access the zeroing screen as show below. During the zeroing, you can select
Cancel to cancel the zeroing.