Alarms and Prompt Messages Alarms and Messages
8 - 20 Operator’s Manual of Anesthesia System
Calibrate
Flow Sensor
and Insp
Valve
1. Calibration table
isn't found in EEPROM.
2. Checksum of the
calibration table does
not match.
1. Calibrate flow
sensors through
Service menu after
entering Standby.
2. Please contact
Mindray Technical
Support if the
problem persists.
Low Startup Manual
Only
Calibrate
Pressure
Sensor and
PEEP Valve
1. Calibration table
isn't found in EEPROM.
2. Checksum of the
calibration table does
not match.
Please contact
Mindray Technical
Support for
pressure
calibration.
Low Startup Manual
Only
Calibrate
21% and
100% O
2
Sensor
1. Calibration table
isn't found in EEPROM.
2. Checksum of the
calibration table does
not match.
1. Calibrate O
2
sensors through
System menu after
entering Standby.
2. Please contact
Mindray Technical
Support if the
problem persists.
Low Startup All
Ventilator
Initialization
Error
After power on, CPU
board cannot send the
parameter settings to
the ventilator control
board.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup Non-
Functional
Ventilator
Initialization:
Time out
Self-test result
cannot be obtained
due to an internal
communication error.
High Startup Non-
Functional
Drive Gas
Pressure Low
Drive Gas Pressure is
low.
1. Connect the
drive gas pipeline.
2. Check that until
drive gas pressure
become normal.
High Startup All
O
2
Supply
Failure
O
2
Supply Failure. 1. Connect the O
2
pipeline.
2. Check that until
O
2
supply pressure
become normal.
High Startup All
Power
Supply
Voltage Error
Power supply voltage
error.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup Manual
Only
RT Clock
Needs
Battery
There is no button
battery cell available in
the system, or the
button battery cell
power is depleted.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup All
RT Clock
Failure
RT chip malfunction. High Startup All
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
STARTUP
RESULT IF
FAIL
Table 8-4 Startup Alarm Messages