Operator’s Manual of Anesthesia System 8 - 21
Alarms and Messages Alarms and Prompt Messages
8.8.2.2 CPU Board Runtime Alarm
Keyboard
Self Test
Error
Keyboard Self Test
Error.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup Non-
Functional
Keyboard
Self Test:
Time out
Keyboard Self Test
result cannot be
obtained due to
communication error.
High Startup Non-
Functional
AG Self Test
Error
AG module self-test
failure.
1. Re-plug or
replace the AG
module.
2. Repeat the test.
3. Please contact
Mindray Technical
Support if the
problem persists.
Low Startup All
AG: Time out External AG self test
result cannot be
obtained due to
communication error.
Low Startup All
Internal AG
Error 02
Internal AG Module
Self Test failure.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Low Startup All
Internal AG:
Time out
Internal AG self test
result cannot be
obtained due to
communication error.
Low Startup All
AGSS Self
Test Error
AGSS self-test failure. 1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Low Startup All
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
STARTUP
RESULT IF
FAIL
Table 8-4 Startup Alarm Messages
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
DISABLE
IN
STANDBY
MODE
IP Address
Conflict
The IP address of the
anesthesia system is
the same as the IP
address of another
device in the local
network.
Check the IP
setting.
Medium Runtime No
Manual Only POST test failed and
the result is [Manual
Only].
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Medium Runtime No
Manual
Only - Leak
Test Fail
Circuit leakage test
failed and the result is
[Manual Only].
Medium Runtime No
Auto
Ventilation
is Non-
Functional
The system's Auto
Ventilation mode is
not available.
High Runtime /
Table 8-5 CPU Board Runtime Alarm Messages

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals