Alarms and Prompt Messages Alarms and Messages
8 - 22 Operator’s Manual of Anesthesia System
8.8.2.3 Power Supply Board Runtime Alarm
NOTE: If the power supply board loses communication with the CPU board for
10 seconds, the alarm buzzer will be turned on.
Status
Screen
Comm Stop
The CPU board lost
communication with
the status screen.
Please contact
Mindray Technical
Support.
High Runtime No
HFNC
Module
Comm Stop
The CPU board lost
communication with
the HFNC module.
Please contact
Mindray Technical
Support.
High Runtime No
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
DISABLE
IN
STANDBY
MODE
Table 8-5 CPU Board Runtime Alarm Messages
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
DISABLE
IN
STANDBY
MODE
Power
System
Comm Stop
Lost communication
with CPU board for
10 seconds.
Please contact
Mindray Technical
Support.
High Runtime No
Power
Supply
Voltage Error
Power supply voltage
error.
High Runtime No
Low Battery
Voltage!
Battery voltage is low. 1. Check the power
supply.
2. Connect AC
power immediately.
High Runtime No
System
going
DOWN,
Battery
depleted!
Battery voltage is too
low.
1. Connect AC
power immediately.
2. In emergency,
please use Manual
ventilation.
High Runtime No
Battery
Undetected
Battery undetected. Please contact
Mindray Technical
Support.
Medium Runtime No
Battery in
Use
AC power fail. Connected to AC
power source.
Low Runtime No
Heating
Module
Failure
1. The temperature
difference between
the two resistances
is large.
2. The temperature of
one of the resistances
is too high.
Please contact
Mindray Technical
Support.
Low Runtime No
Battery
Failure
Battery failure. Please contact
Mindray Technical
Support.
High Runtime No
Table 8-6 Power Supply Board Runtime Alarm Messages