Cleaning and Disinfection Maintenance
9 - 16 Operator’s Manual of Anesthesia System
9.7.3.1.7 Bag Arm
At the base of the bag arm, locate the retaining ring. Turn the ring anticlockwise until it is no longer
threaded.
Figure 9-13 Remove the bag arm
3. Lift the bag arm from the breathing system body.
9.7.3.1.8 Breathing System Body
Hold the handrails on one side of the breathing system body, firmly separate and slide it away from its
mounting arm.
Figure 9-14 Remove the breathing system body
9.7.3.1.9 Inspiration and Expiration Valves
1. Press the buckle to remove the valve dome from the breathing system body.