Operator’s Manual of Anesthesia System 9 - 15
Maintenance Cleaning and Disinfection
Figure 9-10 Remove bypass
9.7.3.1.6 Flow Sensor
1. Loosen the locking nut anticlockwise, remove the inspiration/expiration connector.
Figure 9-11 Remove the inspiration/expiration connector and locking nut
2. Remove the flow sensor from the breathing system.
Figure 9-12 Remove the flow sensor
Buckle button Bypass assembly
Locking nut
Inspiration/
expiration
connector
Flow sensor