Cleaning and Disinfection Maintenance
9 - 14 Operator’s Manual of Anesthesia System
9.7.3.1.4 CO2 Absorbent Canister
1. Rotate the locking mechanism handle clockwise to the unlocked position.
2. Remove the absorber canister.
3. Remove the Pre-Pak or loose fill absorbent from the canisters.
WARNING: Use extreme care while handling the CO2 absorbent as it is a caustic
irritant.
WARNING: Please dispose of discarded absorbent following the requirements of
absorbent manufacturers.
WARNING: To disassemble the bypass assembly, hold the bypass assembly up with
one hand to prevent it from falling and press the buckle button on the
bypass assembly with the other hand.
9.7.3.1.5 Bypass
Press and hold the buckle on the bypass assembly to take out the bypass assembly downward.
Figure 9-8 Canister unlocking Figure 9-9 Canister locking
Unlocking position.
The bypass is turned on.
Gasket
Locking position.
The bypass is turned off.
Gasket