Operator’s Manual of Anesthesia System 9 - 25
Maintenance Cleaning and Disinfection
Figure 9-30 Open the O
2
sensor cover
2. Rotate the O
2
sensor clockwise to install it to the support.
Figure 9-31 Install the O
2
sensor to the support
3. Insert the O
2
sensor support to the breathing system.
Figure 9-32 Insert the O
2
sensor support to the breathing system
4. Insert the O
2
sensor cable.
Oxygen sensor cover switch
O
2
sensor
O
2
sensor support