Loading...
Constraints Among Ventilation Parameters Parameters and Factory Defaults
B - 14 Operator’s Manual of Anesthesia System
B.12 Constraints Among Ventilation Parameters
VENTILATION
MODES PARAMETERS PARAMETER RELATION FORMULA
VCV RR
4 ≤ RR ≤ 100
Vt
10 ≤ Vt ≤ 2000
Plimit Plimit ≥ PEEP+5
SIMV-VC RR
4 ≤ RR≤ 100
Vt
10 ≤ Vt ≤ 2000
ΔPsupp (from
VCV, PCV-VG)
ΔPsupp ≤ Plimit-PEEP
3 ≤ ΔPsupp ≤ 60
Plimit Plimit ≥ PEEP+5
Plimit ≥ ΔPsupp+PEEP
PCV RR
4 ≤ RR ≤ 100
Pinsp If Pinsp is configured:
Pinsp≥PEEP+5
5≤Pinsp≤90
If ΔPinsp is configured:
ΔPinsp≤90 - PEEP
5≤ΔPinsp≤70

Table of Contents