Operator’s Manual of Anesthesia System B - 15
Parameters and Factory Defaults Constraints Among Ventilation Parameters
PCV-VG RR
4 ≤ RR ≤ 100
5≤ Vt ≤ 2000
Plimit Plimit ≥ PEEP+5
SIMV-PC RR
4 ≤ RR ≤ 100
ΔPsupp 3 ≤ ΔPsupp ≤ 60
Pinsp If Pinsp is configured:
Pinsp≥PEEP+5
5≤Pinsp≤90
If ΔPinsp is configured:
ΔPinsp≤90 - PEEP
5≤ΔPinsp≤70
SIMV-VG RR
4 ≤ RR ≤ 100
5≤ Vt ≤ 2000
ΔPsupp (from
VCV, PCV-VG)
ΔPsupp ≤ Plimit-PEEP
3 ≤ ΔPsupp ≤ 60
Plimit Plimit ≥ PEEP+5
Plimit ≥
ΔPsupp+PEEP
VENTILATION
MODES PARAMETERS PARAMETER RELATION FORMULA