Operator’s Manual of Anesthesia System E - 1
E.0
Electromagnetic Compatibility
A8 anesthesia system meets the requirements of IEC 60601-1-2: 2014.
NOTE: The A8 anesthesia system needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC
information provided below.
NOTE: Use of portable or mobile communications devices will degrade the
performance of A8 anesthesia system.
NOTE: A8 anesthesia system is intended for use in professional healthcare
facility environment. If it is used in special environment, such as
magnetic resonance imaging environment, A8 anesthesia system may
be disrupted by the operation of nearby equipment.
WARNING: Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic
immunity of A8 anesthesia system and result in improper operation.
WARNING: Use of A8 anesthesia system adjacent to or stacked with other device
should be avoided because it could result in improper operation. If such
use is necessary, A8 anesthesia system and the other device should be
observed to verify that they are operating normally.
WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the A8 anesthesia system, including
cables specified by Shenzhen Mindray Bio-Medical. Otherwise,
degradation of the performance of this device could result.
WARNING: Other devices may interfere with A8 anesthesia system even though
they meet the requirements of CISPR.
WARNING: When the input signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.