Using the QC Programs
8-7
z When switching from the Predilute mode to the Whole Blood mode, the
analyzer will automatically flush the fluidic system.
z Be sure to keep the sample probe tip away from the tube bottom, otherwise
the aspiration volume may be inaccurate.
z When the aspiration is done, remove the sample tube only when the sample
probe is out of the tube.
Running the controls
1. Be sure the System Status area displays “Ready“ and Count Mode area displays
Whole“.
2. Present a vial of control to the sample probe so that the tip is well into the vial, and press
the aspirate key. The System Status area will display “Running” and the analyzer will
start aspirating sample.
3. When you hear the beep and the sample probe is out of the vial, remove the vial. The
sample probe will retract into the analyzer and the analysis progress will be displayed on
the screen.
4. When the analysis is finished, the result will be displayed on the screen and the
NO./Total” in the upper left corner of the screen will automatically increase by 1 and the
sample probe will be replaced.
z If the analyzer detects WBC/RBC clogging or bubbles during the analysis,
the corresponding error messages will be displayed in the upper left corner
of the screen and the results of all the related parameters will be invalidated.
See Chapter 11 Troubleshooting for solutions.
z If the ambient temperature is outside the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable. See Chapter 11 Troubleshooting for solutions.
Browsing results of other L-J analyses
To browse the result of the preceding or following L-J analysis, press [PgUp] or [PgDn].