Using the QC Programs
8-18
saved only when both the expected result and limit are valid.
Figure8-21 AnInvalid datemessage box
If all the entries are correct, a message box will pop up to remind you to save the changes, as
Figure 8-22 shows. CLICK
Enter” to save the changes and exit to the system menu (or the
“Count” screen); CLICKCancel” to abort the changes and exit to the system menu (or the
Count” screen).
Figure 8-22 A message box to confirm the changes
8.3.2 Running the Controls
Selecting Whole Blood mode
Press [MENU] and SELECTMode” to enter the “Mode”screen. SELECT Whole Blood
from the“Sample Mode” pull-down list.
Entering the “
X
Count” screen
Press [MENU] to enter the system menu. SELECT Quality Control
X
Analysis
X
Count File 1” to enter the “
X
Count”screen, as Figure8-23 shows.