Using the QC Programs
8-23
The 12 parameters are displayed on three screens, 4 parameters on every screen, as
Figure8-27 to Figure 8-29 show. The saved
X
analysis results are sequentially displayed in
the
X
graph, the latest on the utmost left (No.1).
The
X
graph can be interpreted as follows:
The x-coordinate represents the number of the
X
analyses performed; the y-coordinate
represents the results of the
X
analyses;
For every parameter, its
X
graph can display maximum 31 points;
For every parameter, the upper dash line represents the expected result + limit;
For every parameter, the lower dash line represents the expected result – limit;
For every parameter(e.g. WBC), the three numbers to the left of the graph are:
10.5 – the expected result limit;
10.0 – the expected result;
9.5 – the expected result – limit.