Using the Calibration Programs
9-19
Figure9-19 “Sample Table Review”screen
Check the reproducibility as instructed in Chapter 7.11. If the reproducibility meets the
requirements listed in Table 9-1, record the Mean of the 10 runs. If the means of any
parameter falls outside the expected range (see the instructions for use of the calibrator),
calibrate the analyzer as instructed below, otherwise the calibration is not necessary.
If the reproducibility of the calibrated parameter does not meet the requirements of Table 9-1,
you must try to find out the reason and re-run the calibrator after you have solved the problem.
If necessary, contact Mindray customer service department or your local distributor for
assistance.
It is recommended that you create a log table for your analyzer. This log table should contain
all necessary information that is pertinent to your analyzer. Suggested items that you may
want to include in the log table are:
Calibration date
Supplier of calibrator
Lot number
Expected results and limits
Result of background check.
Enter the administrator password instructed in Chapter 5.4.1 and then choose one or several
parameters among WBC, RBC, HGB, MCV and PLT for calibration.