Maintaining Your Analyzer
10-11
3. When the screen reminds you for the second aspiration, present the cleanser to the
probe again and press [ENTER]. When you hear the beep and the sample probe is out
of the bottle, remove the cleanser and the priming progress is displayed on the screen,
as Figure10-9 shows.
Figure10-9 Priming baths and fluidic lines again
4. The cleaning process will last about 15 minutes and you may press [ENTER] to stop it
any time. Note that a shortened priming process may not be as effective as a complete
one.
5. When the cleaning is done, press [ENTER] to flush the bath and tubing, after which
screen will display “Probe Cleanser Cleaning End”.