Maintaining Your Analyzer
10-12
Figure10-10 Cleaning process
To make sure this analyzer functions normally, every time the accumulated analyzed whole
blood samples reach 300 or prediluted blood samples reach 150, a message box will pop up
to remind you to perform the “Probe cleanser cleaning” procedure, as Figure10-11 shows.
CLICK “Enter” to proceed with the cleaning; CLICK “Cancel” to cancel the cleaning.
Figure10-11 A message box to confirm the cleaning
10.3.7 E-Z Cleanser Cleaning
z Samples, controls, calibrators and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
You can use the E-Z cleanser, an enzyme based, isotonic cleaning solution and wetting agent,
to clean the tubing and bath by performing the “E-Z Cleanser Cleaning” procedure.
Follow the steps given below to perform the procedure: