Maintaining Your Analyzer
10-13
1 At the “Maintenance” screen, SELECTE-Z Cleanser Cleaning”;
2 Present the cleanser to the probe and press [ENTER] to aspirate the cleanser.
When you hear the beep and the sample probe is out of the bottle, remove the
cleanser. This analyzer will automatically prime the baths and fluidic lines with the
aspirated cleanser and the progress is displayed on the screen, as Figure10-12
shows;
Figure10-12 Priming the baths and fluidic lines
3 When the priming is done, the cleaning process begins, as Figure10-13 shows. The
default cleansing time is 8 hours and you may press [ENTER] to stop the process
any time;
4 When the cleaning is done, press [ENTER] to drain the baths and fluidic lines, as
Figure10-14 shows. When the draining is done, the screen will display “E-Z
Cleanser Cleaning End”.