Maintaining Your Analyzer
10-20
10.3.11 Empty Tubing
z Samples, controls, calibrators and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z Be sure to do the “Empty Tubing” procedure before relocating the analyzer.
If this analyzer is not to be used for a long time or it is to be maintained, be sure to perform
the “Empty Tubing” procedure to drain the fluidic lines.
Follow the steps given below to do so:
1. At the “Maintenance” screen, press the appropriate arrow keys ([←][→] [↑][↓]) to move
the cursor to “Empty Tubing”.
2. Follow the displayed instructions to remove the diluent, rinse and lyse pickup tubes from
this analyzer and then press [ENTER] to start the draining process, Figure10-22.
Figure10-22 Draining the fluidic lines
3. When the draining is done, the screen will display “Turn off this analyzer” and you
should turn off this analyzer as instructed.