Maintaining Your Analyzer
10-21
10.3.12 Wipe Block Cleaning
After being used for a long time, the bottom of the probe wipe may be contaminated by blood
and the inside of wipe may also be contaminated by the dirt sucked in. So you need to clean
the probe wipe regularly.
Follow the steps given below to do so:
1. At the “Maintenance” screen, SELECTClean wipe block”.
2. Present the probe cleanser to the sample probe and press [ENTER] to aspirate the
cleanser. When you hear the beep and the sample probe is out of the bottle, remove the
cleanser.
3. Push the right door latch in the direction indicated in Figure10-23.
Figure10-23
4. Lift up the front panel latch as indicated in Figure10-24 and open the front panel.
Figure10-24