Troubleshooting Your Analyzer
11-4
Pressure 1 low
The pressure inside the
vacuum chamber does not
reach the expected value
within the given time.
1. Enter the “Service → System Test”
screen and do the “Pressure 1”
procedure as instructed in Chapter
10.6. The error will be removed if the
test result is normal;
2. If the problem remains, contact
Mindray customer service
departmentor your local distributor.
Vacuum Filter
Error
The air inside the vacuum
chamber is not extracted
within the given time.
1. Enter the “Service → System Test”
screen and test the “Vacuum” as
instructed in Chapter 10.6. The error
will be removed if the test result is
normal;
2. If the problem remains, change a
new filter;
3. If the problem remains, contact
Mindray customer service
department or your local distributor.
11.3.2 Reagent errors
Error Message Possible Cause(s) Recommended Action
Lyse Empty
No lyse or a malfunctioning
level transducer.
1. Check if the lyse has run out and if
so;
2. Change a new container of lyse as
instructed in Chapter 4.4.2;
3. If the problem remains, contact
Mindray customer service
department or your local distributor.
Diluent Empty No diluent or a malfunctioning
level transducer.
1. Check if the lyse has run out, and if
so;
2. Change a new container of diluent