Troubleshooting Your Analyzer
11-5
as instructed by Chapter 4.4.2;
3. If the problem remains, contact
Mindray customer service
departmentor your local distributor.
Rinse Empty No rinse or a malfunctioning
level transducer.
1. Check if the rinse has run out, and if
so,
2. Change a new container of rinse as
instructed by Chapter 4.4.2;
3. If the problem remains, contact
Mindray customer service
department or your local distributor.
Rinse Expiry Expired rinse or wrong
expiration setting
1. Check if the rinse has expired. If so,
change a new container of rinse as
instructed by Chapter 4.4.2;
2. If not, reset the expiration date as
instructed in Chapter 5.10.1.
Diluent Expiry Expired diluent or wrong
expiration setting
1. Check if the diluent has expired. If
so, change a new container of
diluent as instructed by Chapter
4.4.2;
2. If not, reset the expiration date as
instructed in Chapter 5.10.1.
Lyse Expiry
Expired lyse or wrong
expiration setting
1. Check if the lyse has expired. If so,
change a new container of lyse as
instructed by Chapter 4.4.2;
2. If not, reset the expiration date as
instructed in Chapter 5.10.1.