Troubleshooting Your Analyzer
11-8
Chapter 10.6;
2. The error will be removed if the test
result is normal;
3. If the problem remains, contact
Mindray customer service
department or your local distributor.
PLT Interrupt Error Something is wrong with the
A/D part of the CPU board.
1.
Enter the “Service System
Test” screen and check the PLT
AD interrupt as instructed in
Chapter 10.5
;
2. The error will be removed if the test
result is normal;
3. If the problem remains, contact
Mindray customer service
department or your local distributor.
11.3.4 Power supply errors
Error Message Possible Cause(s) Recommended Action
DC/DC Error Something is wrong with the
internal DC power supplies.
1. Enter the “Service System
Status” screen and record the
DC-DC 12V” and “DC-DC -12V
values;
2. Shut down the analyzer and
contact Mindray customer service
department or your local distributor.
5V Power Error
Something is wrong with the
power board.
1. Enter the “Service System
Status” screen and record the “5V
voltage;
2. Shut down the analyzer and
contact Mindray customer service
department or your local distributor.
3.3V Power Error
Something is wrong with the
5V power supply.
1. Enter the “Service System
Status” screen and record the