Appendices
B-3
B.5 Performance specifications
B.5.1 Operating range
Parameter Operating range
WBC (10
9
/L) 0.0 - 999.9
RBC (10
12
/L) 0.00 - 9.99
HGB (g/L) 0 - 300
MCV (fL) 0.0 - 250.0
PLT (10
9
/L) 0 - 2999
B.5.2 Normal background
Parameter Background result
WBC
0.3 × 10
9
/ L
RBC
0.03× 10
12
/ L
HGB 1 g / L
HCT 0.5 %
PLT
10 × 10
9
/ L
B.5.3 Linearity range
Parameter Linearity range
WBC (10
9
/L) 0.3 - 99.9
RBC (10
12
/L) 0.20 - 8.00
HGB (g/L) 10 - 250
PLT (10
9
/L) 10 - 999