Appendices
B-4
B.5.4 Reproducibility
These reproducibility requirements applies only to the situation in which 11 normal-level
controls have been run and the results of the 2
nd
to 11
th
runs are used to calculate the
reproducibilities.
Parameter Condition
Reproducibility(CV%)
WBC
7.0 - 15.0 × 10
9
/ L
≤ 2.5
RBC
3.50 - 6.00 × 10
12
/ L
≤ 2.0
HGB 110 - 180 g/L ≤ 1.5
MCV 80.0 - 110.0 fL ≤ 0.5
PLT
150 - 500 × 10
9
/ L
≤ 5.0
B.5.5 Carryover
Parameter Carryover
WBC ≤ 0.5 %
RBC ≤ 0.5 %
HGB ≤ 0.5 %
PLT ≤ 1 %
B.6 Input/output device
z Be sure to use the specified devices only.
B.6.1 Display
Color LCD, 10.2″, 800×600.
B.6.2 Keypad
23-key keypad.
B.6.3 Keyboard
PS/2 keyboard.