Appendices
C-4
C.4 Notes
z This equipment must be operated by skilled/trained medical professionals.
z Be sure to operate the analyzer strictly as instructed in this manual.
z This analyzer adopts a fixed decimal point. You can enter the digits without
bothering to look for the [.] on the external keyboard.
z The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
z Store and use the reagents as instructed by instructions for use of the
reagents
z When you have changed the diluent, rinse or lyse, run a background to
ensure that the system is primed immediately prior to running any samples.
z Pay attention to and record the expiration date and open-container stability
of all the reagents. Be sure not to use expired reagents
z Keep the reagents still for a while before using them.
z Be sure the analyzer is off before switching on the power.
z Be sure to retain the shipping carton and all the packing materials, as they
can be used for packaging if analyzer must be reshipped.
z Be sure to place the analyzer on a countertop.
z Be sure to sue the manufacturer-specified reagents, controls and
calibrators.
z After connecting the reagents, be sure to tighten the cap to prevent
contamination.
z Pay attention to and record the expiration dates and open-container stability
days of all the reagents. Be sure not to use expired reagents
z For any reagent, the entered expiration date should be either the expiration
date printed on the labeling or the open-container expiration date, whichever
is earlier. The open-container expiration date is calculated as follows: the
date that container is opened + the open-container stability days.
z For the whole blood samples to be used for WBC differential or PLT count,
you shall store them at the room temperature and run them within 8 hours
after collection.
z Be sure to use clean K
2
EDTA anticoagulant collection tubes, fused silica
glass/plastic test tubes and 20µL borosilicate
glass capillary tubes.