Installing Your Analyzer
4-10
3. Take out the lyse pickup tube with an orange connector (see Figure4-7) from the
accessory box. Take out the lyse container, in which there should be enough lyse.
Remove the container cap and insert the double-pronged end of the tube into the
container and turn (clockwise) the tube’s cap onto the lyse container until properly
secured;
Figure 4-7
4. Place the lyse container onto the shelf and connect the black connector on the cap to the
black fitting and the orange connector to the orange fitting, as Figure 4-8 shows.
Figure4-8