Installing Your Analyzer
4-11
Connecting diluent container
1. Take out the diluent pickup tube with a green connector (Figure 4-9) from the accessory
box;
2. Take out the diluent container, in which there should be enough diluent, and place it on or
below the countertop;
3. Remove the container cap and insert the double-pronged end of the tube into the diluent
container and turn the tube’s cap onto the diluent container until properly secured;
Figure4-9
4. Locate the green fitting, marked DILUENT, in the lower right corner of the back of the
analyzer. Plug the green connector of the tube into the fitting and turn it clockwise until
properly secured;
5. Locate the transducer fitting, marked DILUENT, in the lower right corner of the back of
the analyzer. Connect the wire by pushing in and turning until properly secured.
Connecting rinse container
1. Take out the diluent pickup tube with a blue connector (see Figure 4-10) from the
accessory box;
2. Take out the rinse container, in which there should be enough diluent, and place it on or
below the countertop;
3. Remove the container cap and insert the double-pronged end of the tube into the rinse
container and turn clockwise the tube’s cap onto the diluent container until properly
secured;