Customizing the Analyzer Software
5-22
5.7.2 Setting WBC channel gain
You can adjust the shape of the WBC histogram by adjusting the gain of the WBC channel.
When WBC histograms of most samples are similar to Figure 5-28, it implies too small a
WBC gain and you need to increase the gain appropriately.
Figure 5-28 WBC gain too small
When WBC histograms of most samples are similar to Figure 5-29, it implies too large a
WBC gain and you need to decrease the gain appropriately.
Figure 5-29 WBC gain too large
To increase (or decrease) the gain
1. Enter the administrator password as introduced in Chapter 5.4.1.
2. Enter the ”Gain” screen and ENTER the desired gain into the “WBC (Whole), as Figure
5-30 shows, or “WBC (Predilute)”, as Figure 5-31 shows.