Customizing the Analyzer Software
5-23
Figure 5-30 Setting WBC(Whole)gain
Figure 5-31 Setting WBC(Predilute)gain
5.7.3 Setting the RBC gain
If the MCV results of most calibration or QC runs deviate from the expected result by 6%, you
need to follow the rule below to change the RBC gain to adjust the MCV results.
Assume the expected MCV result is 90.0fL and the obtained MCV result is 82.0fL.
Then