Customizing the Analyzer Software
5-24
1. Enter the administrator password as introduced in Chapter 5.4.1.
2. Enter the “Gain” screen and ENTER a number into the ”RBC” box, as Figure 5-32 shows,
so that RBC “Factor” is as close to 109.8% as possible.
Figure 5-32 Setting RBC gain
5.7.4 Setting HGB channel gain
You can adjust the HGB blank voltage by adjusting the HGB gain. Normally the HGB blank
voltage should be within 3.4 - 4.8 V (4.5V is recommended). To set the HGB channel gain,
1. Enter the administrator password as introduced in Chapter 5.4.1.
2. Enter the Gain” screen and ENTER the desired gain into the ”HGB” box so that the HGB
voltage falls between 3.4 - 4.8 V, as Figure 5-33 shows.
%8.109%100
0.82
0.90
%100
ActualMCV
VExpectedMC
=××