1 - 1
1 Safety
1.1 Safety Information
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could
result in death, serious injury or damage to product/property.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could
result in minor personal injury, product malfunction or damage to product/
property.
NOTE
Provides application tips or other useful information to ensure that you get
the most out of the product.
1.1.1 Warnings
WARNING
To avoid risk of electric shock, the equipment must only be connected to
mains power with protective earth. If a protective earth conductor is not
provided, operate it on battery power, if possible.
To avoid explosion hazard, do not use the equipment in the presence of
oxygen-rich atmospheres, flammable anesthetics, or other flammable
agents.
The equipment is not intended to be used within the Magnetic Resonance
(MR) environment.
Do not use the multiple portable socket outlets (MPSO) or AC mains
extension cords. Ensure that the sum of the individual ground leakage
currents does not exceed the allowable limits.
Do not open the equipment housings. All servicing and future upgrades must
be carried out by trained and authorized personnel.