Mindray BeneFusion eSP Vet - Viewing the Version Information; The Alarm Settings; The Bolus Volume Unit Settings; The Bolus Limit Settings

Mindray BeneFusion eSP Vet
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 - 5
11.10 Viewing the Version Information
11.11 The Alarm Settings
11.12 The Bolus Volume Unit Settings
11.13 The Bolus Limit Settings
NOTE
The range of auto bolus volume can be expanded. Contact our service
personnel to configure the range if needed.
Menu Item Default Setting Function
Version
Information
/ Displays software version, compile time, driver
software, power software, algorithm, etc.
Menu Item Default Setting Function
Alarm Sound
Sound2
Sets the alarm sound mode.
CMS/eGW
Disconnected
Alarm
Off Sets whether the disconnection alarm will be
triggered when the pump is disconnected from
the CMS or eGateway.
Menu Item Default Setting Function
Bolus Volume
Unit
ml Sets the unit of bolus volume.
Menu Item Default Setting Function
Auto Maximum
volume of syringe
Sets the upper limit of the auto bolus volume
setting. If the set bolus volume exceeds the limit,
the pump prompts you to reconfigure the bolus
volume. The setting range is 0.01 ml to the
maximum volume of current syringe.
Manual 3ml Sets the maximum volume of a manual bolus
infusion. The manual bolus infusion stops when
the set volume is reached. The setting range is 1
ml to 20 ml.