A - 9
A.10.2 Infusion Accuracy at 5ml/h
Test conditions:
Syringe brand: B.Braun Original Perfusor Syringe, B.Braun extension set
Syringe size: 50ml
Test interval: t =0.5 minute
WARNING
Infusion accuracy may be influenced by the pump's environment (such as
pressure, temperature, humidity, and any infusion consumables used).
Observation window(min)
Percentage error of flow
Trumpet Curve at 5ml/h (2nd hour)
Overall percentage error
Ep (Min)
Ep(Max)
Rate (ml/h)
Time(min)
Start-up graph first 2 hours at 5ml/h