8 - 3
• If you do not configure the loading dose parameters, the pump infuses at the
Primary Rate until the set VTBI is finished.
8.4 Micro-infusion Mode
Micro-infusion mode is typically use for low rate infusions for patients.
Micro-infusion mode offers three parameters: rate, time and VTBI. When two of these
parameters are entered, the third is calculated by the pump.
The setting ranges of the parameters in micro-infusion mode are as follows:
• Rate setting is necessary for an infusion, while time and VTBI are optional in
the micro-infusion mode.
8.5 Sequential Mode
In sequential mode, you can set several parameter groups. Each group defines a set of
parameters: rate, time and VTBI. The pump infuses at the set sequence.
8.5.1 Adding/Deleting Sequences
You can add up to eleven sequences in the sequential mode. To add or delete a
sequence, follow this procedure:
Parameters Range
Rate
1ml syringe: 0.01 to 50ml/h
2/3ml and 5/6ml syringe: 0.01 to 100ml/h
Syringe of other sizes: 0.1 to 100ml/h
VTBI 0.01 to 1000ml
Rate
Rate
S2 Time
S1 Time
S2 Rate
S1 Rate
S3 Time
S4 Time
S3 Rate
S4 Rate