8 - 4
1. Select a sequence (such as S1) from the parameter setup screen.
2. In the popup dialog, make the following settings:
◆ Select Add Sequence Upward to add a sequence before the current
sequence.
◆ Select Add Sequence Backward to add a sequence after the current
sequence.
◆ Select Delete to delete the current sequence.
8.5.2 Changing the Infusion Parameters
You can change the rate of the current sequence during an infusion. If you want to
change the time or VTBI of the current sequence, press to pause the infusion and
select the desired parameter area to make the change.
To change parameters of other sequences, follow this procedure:
1. Press to pause the infusion.
2. Select .
3. Select the desired parameter area to make the change.
8.6 Intermittent Mode
In the intermittent mode, intermittent infusion and maintenance are performed
alternately and circularly.
■ Intermittent stage: the pump runs the high rate infusion at the set Rate and Intmt.
Vol.
■ Maintenance stage: the pump runs the low rate infusion at the set Maintain Rate
and
Intmt. Time. The pump does not infuse at this stage if the Maintain Rate is not
set.
Time
Rate
Total VTBI
Intmt.
Time
Intermittent
stage
Intmt. Vol.
Maintenance
stage
Maintain Rate