4 - 7
4.6.4 Setting the Bolus Volume Unit
To set the bolus volume, follow this procedure:
1. Swipe the touchscreen from top down select Menu select User
Maintenance input the required password select .
2. Select the Bolus Volume Unit:
ml: the
BolusVTBI unit is ml in each infusion mode.
Dose: in the Dose Mode or Dose Time Mode, Drug Amt. and Volume, or
Conc. is set up, the
BolusVTBI unit is the Drug Amt. unit or the
corresponding unit of the Conc.
4.7 Changing the Infusion Parameters
You can modify rate, dose rate or drug name without stopping the infusion. This
function is called titration.
1. Select the above parameters in the infusion running screen.
2. Reconfigure the parameters in the popup dialogs.
To change other infusion parameters, follow this procedure:
1. Press to pause the infusion.
2. Select the desired parameter area, and reconfigure parameters as per the
prescriber’s order.
3. Select OK to confirm the changing.
4. Press again to resume the infusion.
4.8 Pausing the Infusion
Press to temporarily stops a running infusion.
Press again to restart the infusion after the infusion solution change.
4.9 Setting Keep Vein Open (KVO) Rate
At the end of infusion, the pump continues to infuse at a very low rate. KVO is used to
keep the patient’s vein open, to prevent back flow or vascular occlusion.
The default KVO rate is 0.5 ml/h. To edit the KVO rate, follow this procedure:
1. Swipe the touchscreen from top down select Menu select General Option.
2. Set the KVO Rate. If KVO Rate is zero, the pump will not initiate a KVO infusion
when the preset volume is complete.