A.9 Occlusion Alarm Delay and Bolus Volume
Test conditions:
Syringe brand: B.Braun Original Perfusor Syringe, B. Braun extension line
Test temperature: 20°C ±2°C
WARNING
Occlusion alarm pressure, alarm delays and bolus volume may vary
depending on test conditions, temperature and tube length. Using syringe of
a larger size and infusing at a lower rate may cause longer occlusion alarm
delay.
Syringe size (ml) Rate (ml/h)
Occlusion alarm delay time (hh: mm: ss)
High occlusion alarm
pressure level
Low occlusion alarm
pressure level
20 1
00:33:12 00:02:12
5
00:06:54 00:00:45
50 1
01:45:57 00:06:53
5
00:25:15 00:01:42
Syringe size (ml) Rate (ml/h)
Bolus volume after occlusion (ml)
High occlusion alarm
pressure level
Low occlusion alarm
pressure level
20 5
0.2 0.1
50 5
0.3 0.15