3 - 2
NOTE
Save the packing case and packaging material as they can be used if the
equipment must be reshipped.
3.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in
this manual.
The environment where the equipment is used shall be reasonably free from noises,
vibration, dust, corrosive, flammable, and explosive substances. Moreover, to maintain
good ventilation, the equipment shall be at least 2 inches (5cm) away from around the
cabinet.
When the equipment is moved from one place to another, condensation may occur as a
result of temperature or humidity differences. In this case, never start the system before
the condensation evaporates.
CAUTION
Make sure that the equipment operating environment meets the specific
requirements. Otherwise unexpected consequences, e.g. damage to the
equipment, could result.
3.3 Installation
3.3.1 Pole Clamp Installation
The pole clamp secures the pump to either a horizontal or vertical bar of the medical
supply unit or IV pole. For detailed information on how to install the pole clamp, see The
Pole Clamp Installation Guide.
3.3.2 Stack Rack Installation
Use a stack rack for pump transport or for stacking several pumps together. For detailed
instructions on stack rack installation, see The Stack Rack Installation Guide.
NOTE
Check the medical supply unit and IV pole for stability before mounting the
pumps.
Install a single pole clamp to each pump before mounting the stacked pumps
to the medical supply unit or IV pole.