A - 4
A.6 Wireless Network
A.7 Infusion Specifications
Standards IEEE 802.11a/b/g/n
Modulation mode BPSK,QPSK, QAM
Operating frequency 2.412GHz to 2.484GHz
5.18GHz to 5.24GHz
5.745GHz to 5.825GHz
Data rate IEEE 802.11a: 6 to 54 Mbps
IEEE 802.11b: 1 to 11 Mbps
IEEE 802.11g: 6 to 54 Mbps
IEEE 802.11n: 6.5 to 65 Mbps
Transfer power < 20 dBm (CE requirement: detection mode – RMS)
< 30 dBm (FCC requirement: detection mode – PEAK)
Operating mode Transmitting data through the wireless access point
(AP)
Data security Standard: WPA-PSK and WPA2-PSK
Encryption: TKIP and AES
System capacity
Number of the pumps supported by a single AP: 16
Data transmission delay
between the pump and the
CMS
Total data transmission delay time between the pump
and the CMS is 8s
Interruption number and time
between the pump and the
CMS
Total interruption duration 0.01* total
communication time
(Test within 24 hours, with 16 pumps, in which three
pumps are roaming for 30 times)
Delay time of network
disconnection alarm
14 s
Compatible syringe sizes 1ml, 2ml, 3ml, 5ml/6ml, 10ml/12ml, 20ml, 30ml/35ml,
50ml/60ml (1ml is optional)

Table of Contents