4 - 6
• The delivered bolus volume will be added to the total infusion volume and
subtracted from the volume to be infused (VTBI).
• The pump gives a beep every time a 0.5 ml bolus volume is infused.
4.6.1 Setting the Bolus Rate
To set the bolus rate, follow this procedure:
1. Swipe the touchscreen from top down → select Menu → select General Option.
2. Set the BolusRate.
4.6.2 Automatic Bolus Infusion
To perform an automatic bolus infusion, follow this procedure:
1. Select from the main screen.
2. Set the bolus volume in the popup dialog.
3. Select to start a bolus infusion.
The pump continues the infusion when the configured bolus volume has been infused.
If required, select to stop the bolus infusion.
• If required, adjust the bolus rate in the BolusRate area during an automatic
bolus infusion.
4.6.3 Manual Bolus Infusion
To perform a manual bolus infusion, follow this procedure:
1. Select from the main screen.
2. Set the bolus volume in the popup dialog.
3. Press and hold to deliver the required bolus.
4. Release when the desired bolus volume has been delivered or the bolus
volume limit is reached.
• The manual bolus volume limit is set in the User Maintenance menu. See
11.13 The Bolus Limit Settings.