6 - 5
◆ Continuous: the nurse call signal lasts until the alarm ends. That is to say the
duration of a nurse call signal is equal to that of the alarm condition.
4. Select Trigger Type to set the work mode of the nurse call relay.
5. Select Alarm Level to set the priority of alarms sent to the nurse call system.
6.7 Alarm Solutions
• When an alarm occurs, check the pump’s status and handle the alarm as soon
as possible. If the alarms do not conform with the actual situation, contact
your service personnel.
Alarm Priority Causes Solutions
Occlusion High An occlusion occurred
and the preset pressure
limit is exceeded.
• Check that tubing is not
kink or damaged.
• Check the pressure limit
setting. Increase the limit if
necessary.
Syringe
Empty
High No fluid is left in the
syringe.
• Press to clear the
alarm.
• End the infusion or replace
the syringe.
Syringe
Disengaged
High The syringe is
disengaged.
Reload the syringe.
No Syringe High The syringe is not loaded
properly.
Reload the syringe.
Plugger
Grippers Error
High The plunger grippers do
not work properly.
• Check that the plunger
grippers are not blocked .
• Manually open or close the
plunger grippers.
• If the alarm persists, contact
your service personnel.
Extension
Line
Detached
Low The extension set is
disengaged.
Check and reconnect the
extension set.