13-1
13 Monitoring Temp
13.1 Introduction
You can simultaneously monitor two temperature sites using the iMEC monitors.
13.2 Safety
WARNING
Verify that the probe detection program works correctly before monitoring. Unplug the temperature probe
cable from the T1 or T2 connector, and the monitor can display the message [T1 Sensor Off] or [T2 Sensor
Off] and give alarm tones correctly.
13.3 Making a Temp Measurement
1. Select an appropriate probe for your patient.
2. If you are using a disposable probe, connect the probe to the temperature cable.
3 Plug the probe or temperature cable to the temperature connector.
4. Attach the probe to the patient correctly.
5. Check that the alarm settings are appropriate for this patient.