EasyManua.ls Logo

Mindray iMEC8 Vet - 15 Monitoring Carbon Dioxide (CO ); Introduction

Mindray iMEC8 Vet
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15-1
15 Monitoring Carbon Dioxide (CO
2
)
15.1 Introduction
CO
2
monitoring is a continuous, non-invasive technique for determining the concentration of CO
2
in the patients airway
by measuring the absorption of infrared (IR) light of specific wavelengths. The CO
2
has its own absorption characteristic
and the amount of light passing the gas probe depends on the concentration of the measured CO
2.
When a specific band
of IR light is passed through respiratory gas samples, some of the IR light will be absorbed by the CO
2
molecules. The
amount of IR light transmitted after it has been passed through the respiratory gas sample is measured with a
photodetector. From the amount of IR light measured, the concentration of CO
2
is calculated.
There are two methods for measuring CO
2
in the patients airway:
1. Mainstream measurement uses a CO
2
sensor attached to an airway adapter directly inserted into the patient’s
breathing system.
2. Sidestream/Microstream measurement samples expired patient gas at a constant sample flow from the patient’s
airway and analyzes it with a CO
2
sensor built into the monitor.
The measurement provides:
1. A CO
2
waveform
2. End tidal CO
2
value (EtCO
2
): the CO
2
value measured at the end of the expiration phase.
3. Fraction of inspired CO
2
(FiCO
2
): the smallest CO
2
value measured during inspiration.
4. Airway respiration rate (awRR): the number of breaths per minute, calculated from the CO
2
waveform.

Table of Contents

Related product manuals