3-2
WARNING
When disposing of the packaging material, be sure to observe the applicable waste control regulations and
keep it out of children’s reach.
Before use, please verify that the packages are intact, especially the packages of single use accessories. In
case of any damage, do not apply it to patients.
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual.
The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive,
flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind
shall be left for convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment
shall be at least 2 inches (5cm) away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result of temperature or
humidity difference. In this case, never start the system before the condensation disappears.
CAUTION
Make sure that the operating environment of the equipment meets the specific requirements. Otherwise
unexpected consequences, e.g. damage to the equipment, could result.
Put the equipment in a location where you can easily see the screen, access the operating controls, and
disconnect the equipment from AC power.
3.2 Getting Started
3.2.1 Turning Power On
Once the monitor is installed, you can get ready for monitoring:
1. Before you start to make measurements, check the monitor for any mechanical damage and make sure that all
external cables, plug-ins and accessories are properly connected.
2. Plug the power cord into the AC power source. If you run the monitor on battery power, ensure that the battery
is sufficiently charged.
3. Press the power on/off button on the monitor’s front.
WARNING
Do not use the monitor for any monitoring procedure on a patient if you suspect it is not working properly,
or if it is mechanically damaged. Contact your service personnel or Mindray.